QC Rejection Rules (Westagard Rules) in Simplified Table

Internal quality control results are monitored by using Levy Jennings Chart (LJ Chart / Control chart) with applicable westgard rules.  Herewith i provided the five basic  westgard rules in simplified table to make decision on Accepting/Rejecting the QC run. 

Laboratory professionals can download the below  table (download link provided) and displayed as poster in respective lab sections for easy reference.

qcrules

When a rule is violated :

Warning Rule = use another rules to inspect the control points

Rejection Rule = "Out of Control"
  • Stop Testing
  • Search for recent events that could have caused changes
  • Examine environmental conditions.
  • Identify and correct problem
  • Repeat  testing on patient samples and controls
  • Follow manufacturer’s troubleshooting guide
  • Refer to instructions of manufacturers of equipment, reagents or QC / calibrator
  • Do not report patient samples until problem is solved and controls indicate  proper performance

Points of QC Failures : 
  • Instrument Failure:  
    • Check for error messages (printed or displayed). 
    • Refer to the troubleshooting section of the instrument manual. 
  • Reagent, Quality Control Materials, Calibrators: 
    • In correct reconstitution, i.e. wrong or contaminated diluent or use of the wrong volume of diluent. 
    • In correct storage, e.g. left at room temperature for excessive amount of time. – Prepare fresh (allow enough time to dissolve) 
    • NOTE: Bulk liquid reagents may be contaminated. 
  • Human: 
    • Reread the method 
    • Check whether the correct sample volumes or reagents were used.
    •  Make sure that a step in the procedure was not missed. 
    • Were the correct parameters entered?