Following step by step Instructions suggested for verification of new equipment or method (or Reagent) used in the medical lab prior to testing patient sample analysis.
The instruction was provided in simplified manner for better understanding by the Medical laboratory professionals.
5. The samples should be selected to
cover the entire reportable range of the equipment/method.
10.After completion of study (after 5
days) prepare a comparison plot for all the data to assess the range, outliers
and linearity.
Yc
= a + bXc
SE = Yc – Xc
13. Calculate the Random Error (RE) from
the standard deviation (SD) obtained from replicate experiment (Precision
Verification) by using repeated 20 runs of same patient sample (or) Quality
control material
RE
= 3 ´ SD
14. Derive the Calculated Total Error (
TEcalc ) by
TEcalc = SE + RE
15. Method/Equipment performance is
acceptable when this calculated total error (TEcalc ) is less than the allowable total error
(TEa).
16. Refer CLIA guidelines for
proficiency limits and biological variation values for fixing the acceptable
total allowable error (TEa).
Example - Equipment/Test Method Comparison Report Format (Developed in MS-Excel):
Reference Source :
- Basic Method Validation, 3rd Edition, James O. Westgard,PhD
- NABL - 112 Guideline (Specific Criteria for Accreditation of Medical Laboratories)
POSTED BY:
Mr.V.SELVAM, M.Sc (Biochemistry)
Consultant - ISO 9001 & NABL (ISO 15189 & ISO 17025)
Chennai, Tamilnadu, India
✆ +91-9940014466


Very informative sir. Can we plan this comparison study in a single day after collecting 20 samples covering the entire reportable range?
ReplyDeleteyes can be done. Thank you for your feedback.
DeleteThanks very much for sharing
Delete